TECHNICAL DATASHEET
Medical ATX Power Supply
ATX 12 V 1 U Power Supply
MDS-350AD701
Highlights & Features
• ATX 12 V standard function & form factor
• IEC60601-1 3.1 ed.(Medical) & IEC 62368-1 (ITE)
•
•
•
•
approvals
Suitable for type BF patient access applications
Built-in active PFC
Maximum output power of 350 watts
Compliant with IEC 60601-1-2 4th Ed. requirements
Safety Standards
CB Certified for worldwide use
Model Number:
MDS-350AD701 AA
Unit Weight:
1.2 Kg (2.6 lb)
Dimensions (L x W x H): 194 x 100 x 40.5 mm
7.64 x 3.94 x 1.59 inch
General Description
The MDS-350AD701 series of internal ATX power supplies comes with universal AC input from 90 Vac to 264 Vac. Other
features include low leakage, availability of risk management report, and electric shock protection compliant with 2 x
MOPP requirements. The MDS series is certified for EMC standards according to EN/BS EN 55011 for industrial,
scientific and medical (ISM) radio-frequency equipment; and, EN/BS EN 55032 for Industrial Technology Equipment (ITE)
radio-frequency equipment.
The MDS-350AD701 series comes with both Medical and ITE safety approvals, including UL/CE (5000 m) and CB
certification and are fully compliant with RoHS Directive for environmental protection.
Model Information
Table 1 (For Maximum Output Power Limits, see “Output Power” requirements on Page 2)
Output Voltage
Minimum Load (A)
Normal Load (A)
Maximum Load (A)
Load Regulation
Ripple & Noise
+3.3 V
0.5
8
16
+/- 5%
50 mVpp
+5 V
0.5
9
18
+/- 5%
50 mVpp
+5 Vsb
0.0
1.25
2.5
+/- 5%
150 mVpp
+12 V1
1.0
8
16
+/- 5%
120 mVpp
+12 V2
1.0
8
16
+/- 5%
120 mVpp
-12 V
0.0
0.25
0.5
+/- 10%
120 mVpp
-5 V
0.0
0.1
0.2
+/- 10%
50 mVpp
Model Numbering
MDS
350
A
Delta Medical
Power Supply
Max Wattage in
the Product
Series.
Family Code
1
A~ Z
D
Product Type
D: Enclosure
701
Output Voltage
701: +3.3 V, +5 V, +5 Vsb,
+12 V1, +12 V2, -12 V, -5 V
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (June 2022, Rev.07)
A
Revision Code
A
Revision code
TECHNICAL DATASHEET
Medical ATX Power Supply
ATX 12 V 1 U Power Supply
Specifications
Input Ratings / Characteristics
Nominal Input Voltage
100-240 Vac
Input Voltage Range
90-264 Vac
Nominal Input Frequency
50-60 Hz
Input Frequency Range
47-63 Hz
Input Current (max)
6A
Input Surge Voltage (max)
300 Vac for 100 ms
Full load Efficiency (typ.)
82% @ 115 Vac/60 Hz
85% @ 230 Vac/50 Hz
Inrush Current (max)
No damage, cold start
Leakage Current (max)
0.1 mA @ NC, 0.3 mA @ SFC1)
Output-PE (protective earth) Leakage current for BF application (max)
0.1 mA @ NC, 0.5 mA @ SFC
Power Factor (min)
0.90 @ 115 V/50 Hz, 230 V/50 Hz, full load
1) NC: normal condition, SFC: single fault condition
Output Ratings / Characteristics
Nominal Output Voltage (V rated)
+12 V1, +12 V2, +5 V, +3.3 V, +5 Vsb, -5 V, -12 V
Output Voltage Tolerance
See Table 1 on page 1
Output Power
Maximum combined output power of +3.3 V & +5 V
outputs, shall not exceed 130 W.
Maximum combined output power of +12 V1, +12 V2,
+3.3 V & +5 V outputs shall not exceed 331 W
Maximum combined output power from all the outputs
shall not exceed 350 W
±0.5%
Line Regulation (max)
Load Regulation (max)
Ripple & Noise (typ.)
See Table 1
Start-up Time (max) (Remote on_off @ TTL Low, then AC applied)
1000 ms @ 115 Vac
Hold-up Time (min)
17 ms @ 115 Vac / 60 Hz
Dynamic Response (Overshoot & Undershoot O/P Voltage)
± 10% @ 50-100% load1)
Rise time (max)
20 ms
Remote On_Off/INHIBIT
TTL Active Low: Output Enable
TTL Active High: Output Disable
Power Good (PG) Signal(See page 9 for details)
1.
2.
1) When the dynamic load is applied to one of the outputs, other loads is at minimum load
2
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (June 2022, Rev.07)
PG changes to TTL High 100-500 milliseconds after
all outputs reach their final steady state value.
After input AC is removed, PG changes to TTL Low
at least 1 millisecond before any output falls out of
regulation range.
TECHNICAL DATASHEET
Medical ATX Power Supply
ATX 12 V 1 U Power Supply
Ripple & Noise measurement circuit for each output
Mechanical
Case Chassis
SECC (Steel, Electrogalvanized, Cold Rolled, Coil)
Case Cover
SECC (Steel, Electrogalvanized, Cold Rolled, Coil)
Dimensions (L x W x H)
194 x 100 x 40.5 mm (7.64 x 3.94 x 1.59 inch)
Unit Weight (typ.)
1.2 Kg (2.6 pounds)
Indicator
None
Cooling System
Forced Cooling
Environment
Surrounding Air Temperature
Operating
Storage
0°C to +50°C
-40°C to +85°C
Operating Humidity
5-95% RH (Non-Condensing)
Operating Altitude
5,000 meters (16,400 feet)
Shock Test (Non-Operating)
50 G, 11 ms, 3 shocks for each direction
Vibration (Operating)
5-500 Hz, 2 Grms, 15 minute for each three axis
3
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (June 2022, Rev.07)
TECHNICAL DATASHEET
Medical ATX Power Supply
ATX 12 V 1 U Power Supply
Protections
(Latch Mode protection requires removal of fault, plus removal and reapplication of input AC voltage, in order to restart the power
supply. See page 9 for additional details on the different types of protection.)
Overvoltage (max)
For +3.3 V output, Maximum voltage < 4.8 V
For +5 V output, Maximum voltage < 7 V
For +12 V1 & +12 V2 outputs, Maximum voltage < 16V
Protection Method: Latch Mode
For +5 Vsb output, Maximum voltage < 10 V
Protection Method: Latch Mode
Over load / Over current (max)
For +3.3 V output, Maximum current < 25 A
For +5 V output, Maximum current < 25 A
For +12 V1/+12 V2 output, Maximum current < 25 A
Protection Method: Latch Mode
For +5 Vsb output, Maximum current < 5 A
Protection Method: Auto-recovery Mode
Over Temperature
Protection Method: Latch Mode
Short Circuit on any output.
+3.3 V, +5 V, +12 V1 and +12 V2, Protection Method:
Latch Mode
+5 Vsb, Protection Method: Auto-recovery
Reliability Data
MTBF (Minimum) at 115 Vac, 35 °C with 350 W
Forced Air Flow
500 K hrs based on Telecordia SR-332
Operating Life (Minimum) at 115 Vac, 25°C with 350 W
Forced Air Flow
26,280 hrs
Safety Standards / Directives
Medical Safety
IEC 60601-1 Edition 3.1 (2012), EN60601-1 (2006) +
A11 + A1 + A12, CAN/CSA-C22.2 NO. 60601-1:14,
ANSI/AAMI ES60601-1:2005/(R)2012
ITE Safety
IEC 60950-1 (Ed.2,2005) & IEC 62368-1 CB report,
CE
In conformance with EMC Directive 2014/30/EU and
Low Voltage Directive 2014/35/EU
EN 60601-1: 2006 + A11: 2011 + A1: 2013 + A12:
2014 & EN 60601-1-2: 2015
UKCA
In conformance with Electrical Equipment (Safety)
Regulations 2016, and Electromagnetic Compatibility
Regulations 2016,
Medical Devices Regulations 2002(UK MDR 2002)
Galvanic Isolation
Input to/Output(2XMOPP)
4000 Vac
Input to/Ground(1XMOPP)
1500 Vac2)
Output to/Ground(1XMOPP) 1500 Vac (Type BF application rated)
2) PSU can support PoE applications with Primary to FG 2500 Vac test.
4
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (June 2022, Rev.07)
TECHNICAL DATASHEET
Medical ATX Power Supply
ATX 12 V 1 U Power Supply
EMC
EMC / Emissions
Harmonic Current Emissions
EN/BS EN 55011 & Compliant with EN/BS EN 55032,
FCC Title 47:Class B
IEC61000-3-2
Meet Class D limit
Immunity to
Voltage Flicker
IEC61000-3-3
Electrostatic Discharge
IEC61000-4-2
Level 4 Criteria A1)5)
Air Discharge: 15 kV
Contact Discharge: 8 kV
Radiated Field
IEC61000-4-3
Level 3 Criteria A1)
80 MHz-1000 MHz, 10 V/m AM modulation
IEC60601-1-2
Criteria A1)5)
80 MHz-2700 MHz, 10 V/m AM modulation
385 MHz-5785 MHz, 28 V/m Pulse mode and other
modulation
Electrical Fast Transient / Burst
IEC61000-4-4
Level 3 Criteria A1) :2 kV
Surge
IEC61000-4-5
Level 3 Criteria A1)5)
Common Mode3): 2 kV
Differential Mode4): 1 kV
Conducted
IEC61000-4-6
Level 2 Criteria A1)5)
150 kHz-80 MHz, 3 Vrms, 6 Vrms at ISM bands and
Amateur radio bands
Power Frequency Magnetic Fields
IEC61000-4-8
Criteria A1)5)
Magnetic field strength 30 A/m
Voltage Dips5)
IEC60601-1-2
Criteria C2)
30% 500 ms
Criteria A1)
100% 10 ms at step: 45 degree
Criteria C2)
100% 20 ms
Criteria C2)
100% 5000 ms
1) Criteria A: Normal performance within the specification limits
2) Criteria C: Output out of regulation, or shuts down during test. (Need to recycle AC power cord or reset enable signal to normal operation after test)
3) Asymmetrical: Common mode (Line to earth)
4) Symmetrical: Differential mode (Line to line)
5) Compliant with IEC-60601-1-2 4th edition requirements.
5
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (June 2022, Rev.07)
TECHNICAL DATASHEET
Medical ATX Power Supply
ATX 12 V 1 U Power Supply
Dimensions
L x W x H: 194 x 100 x 40.5 mm (7.64 x 3.94 x 1.59 inch)
Mechanical drawing
Notes
•
Dimension are in mm.
•
Built-in cooling fan. Must prevent dust suction into power supply, or use natural convection power supply if any concerns.
6
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (June 2022, Rev.07)
TECHNICAL DATASHEET
Medical ATX Power Supply
ATX 12 V 1 U Power Supply
DC Output Cable
7
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (June 2022, Rev.07)
TECHNICAL DATASHEET
Medical ATX Power Supply
ATX 12 V 1 U Power Supply
Functions
Start-up Time
The time required for the output voltage to reach 90% of its final steady state value, after the input voltage is applied.
Rise Time
The time required for the output voltage to change from 10% to 90% of its final steady state value.
Hold-up Time
Hold-up Time is the time between removal of input AC voltage, and the subsequently falling output to reach 90% of its steady state
value.
8
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (June 2022, Rev.07)
TECHNICAL DATASHEET
Medical ATX Power Supply
ATX 12 V 1 U Power Supply
Dynamic Response
Over Temperature Protection
The power supply output voltage will remain within ±10% of its
steady state value, when subjected to a dynamic load 50 to 100%
of its rated current.
■ 50 to 100% Load
As mentioned above, the power supply also has Over
Temperature Protection (OTP). This is activated when the
overload condition persists for an extended duration and the
output current is below the overload trigger point but >100% load.
In the event of a higher operating temperature condition at 100%
load, the power supply will run into OTP when the surrounding air
temperature is higher than the operating temperature. When
activated, the output voltage will go into latch mode until the input
voltage is removed; then, reapplied, and the surrounding air
temperature drops to its normal operating temperature.
Power Good
Inrush Current
Inrush current is the peak, instantaneous, input current measured
and, occurs when the input voltage is first applied. For AC input
voltages, the maximum peak value of inrush current will occur
during the first half cycle of the applied AC voltage. This peak
value decreases exponentially during subsequent cycles of AC
voltage.
The Power Good signal prevents a computer from attempting to
operate on improper voltages and damaging itself by alerting it to
improper power supply. The ATX specification defines the PowerGood signal as a +5 volt (V) signal generated in the power supply
when it has passed its internal self-tests and the outputs have
stabilized. This normally takes between 0.1 and 0.5 seconds after
the power supply is switched on. The signal is then sent to the
motherboard, where it is received by the processor timer chip that
controls the reset line to the processor. When plug in AC power
cord, +5 Vsb will start first, then Remote On/Off will be pulled low
by computer, +12 V1/+12 V2/+5 V/+3.3 V will power on and
Power good signal will pull high after 100ms~500ms. After input
AC is removed, Power Good signal changes to TTL Low at least
1 millisecond before any output falls out of regulation range.
Graph illustrating the AC input, +5 Vsb, Remote On/Off,
+12 V1/12 V2/5/3.3 V, Power good signal
Overvoltage Protection
The power supply’s overvoltage circuit will be activated when its
internal feedback circuit fails. The output voltage shall not exceed
its specifications defined on Page 4 under “Protections”. Power
supply will latch off, and require removal/re-application of input
AC voltage in order to restart.
Short Circuit Protection
The power supply’s output OLP/OCP function also provides
protection against short circuits. When a short circuit is applied to
any of the outputs, the power supply will latch off.
Overload & Over current Protections
The power supply’s Overload (OLP) and Over current (OCP)
Protections will be activated before the output currents reach the
limits defined on Page 4. Once the power supply has reached
their maximum power limit, the protection will be activated and
the power supply will latch off.
9
Remote On_Off/INHIBIT
Remote ON_OFF/INHIBIT can be used to enable or disable all
outputs, except for the +5 Volt Standby. This signal can be pulled
down to a low level of 0.3 volts, or shorted to DC-Return, in order
for the main output to be enabled; and, floated (no connection to
the signal), or pulled up to a value greater than or equal to 3 volts,
in order to disable the main output. When the main output is
disabled, the +5 V Standby output will continue to operate.
All parameters are specified at 25°C ambient unless otherwise noted.
www.DeltaPSU.com (June 2022, Rev.07)
TECHNICAL DATASHEET
Medical ATX Power Supply
ATX 12 V 1 U Power Supply
Certificate
Delta has been certified as meeting the requirement of ISO 13485: 2003 and EN ISO
13485:2012 for the design and manufacture of switching power supply and adaptor for medical
device.
In addition to a UL Total Certification Program (TCP) approved client laboratory for IEC 62368-1.
Delta also has participated UL Client Test Data Program (CDTP) for IEC 60601.
Attention
Delta provides all information in the datasheets on an “AS IS” basis and does not offer any kind of warranty through the information for
using the product. In the event of any discrepancy between the information in the catalog and datasheets, the datasheets shall prevail
(please refer to www.DeltaPSU.com for the latest datasheets information). Delta shall have no liability of indemnification for any claim
or action arising from any error for the provided information in the datasheets. Customer shall take its responsibility for evaluation of
using the product before placing an order with Delta.
Delta reserves the right to make changes to the information described in the datasheets without notice.
Manufacturer and Authorized Representatives Information
Manufacturer
Thailand
Delta Electronics (Thailand) PCL.
909 Pattana 1 Rd., Muang, Samutprakarn, 10280 Thailand
Taiwan
Delta Electronics, Inc.
3 Tungyuan Road, Chungli Industrial Zone, Taoyuan County
32063, Taiwan
Authorized Representatives
The Netherlands
Delta Greentech (Netherlands) B.V.
Zandsteen 15, 2132 MZ Hoofddorp, The Netherlands
10
United Kingdom
Delta Electronics Europe Limited
1 Redwood Court, Peel Park Campus,
East Kilbride, Glasgow, G74 5PF, United Kingdom
All parameters are specified at 25°C ambient unless otherwise noted.